Description
TEST SPECIMEN FOR COMPUTED RADIOGRAPHY SCANNERS
The CR Phantom can test all the important performance characteristics of a CR scanner system including basic spatial resolution (unsharpness), contrast, MTF, laser beam jitter, scanner slipping and scanner shading. These tests are described in detail in the ASTM E 2445 and EN 14784-1 standards and should be performed periodically to ensure proper and accurate system operation.
- T-target (brass). Laser beam jitter, MTF check, Blooming (Flare)
- Duplex wire-type IQI 15D (hi-res). Basic Spatial Resolution (Unsharpness)
- BAM snail. Central beam alignment
- Line pair IQI type 53-b. Line pair resolution
- Measuring points. Shading correction
- Cassette positioning locator. Positioning of cassette (imaging plate)
- Homogeneous strip (aluminium). Scanner slippage, scanner shading
- Lucite plate. Carrier plate
- cm/inch ruler. Linearity check
- Contrast sensitivity gauge. Contrast sensitivity check
Link: CR Phantom